Potential causes of decreasing ninhydrin sensitivity:
- Low grade nitrogen or inert gas supply (must be 99.999%) used in
preparation or on instrument.
- Incomplete dissolution of reagent before use.
- Nitrogen supply interrupted:
- Cylinder valve stuck
- PTFE purge tube trapped or disconnected
- Reservoir bottle seal leakage
- Reservoir still contaminated with detergent after washing (also gives
yellow or pink colours with Virkon, decon). Use only water for
rinsing reservoir.