Can you tell me the standard procedure in using Internal Standard in the sample preparation?

Answer

To use the internal standard norleucine (Nleu) for calculation make up a stock solution at 50nmol/20µL and add 1 volume of this solution to 4 volumes of your amino acid calibration standard to give a final concentration of 10nmol Nleu/20µL of standard loaded.

For sample preparation, add the Norleucine stock solution to your sulphosalicylic acid deproteinisation solution to give a final loaded amount of 10nmol Nleu/20µL.

To calculate amino acids following preparation and check for sample losses, select the internal standard report option in your method.