Frequently Asked Questions

Microplate Instrumentation

Question

What is the indication range of the Anthos 2010 how does this differ from the linear range?

Answer

The Biochrom Anthos 2010 has an indication range of 0.000-3.300 OD and a linear range of 0.100-2.500 OD.

The Biochrom Anthos 2010 is capable of a high degree of linearity:

<± 0.75% and 0.005 OD

from 0.100-2.500 OD. We recommend that the broader indication range should be used only for relative determinations but not for determining the quantity of the absorbing material. Only measurements made within the linear range should be used for quantitative calculations.


Feedback

0 rating from 0 votes

Your feedback is important to us. Why not take a moment to let us know how helpful you found our answer by using the star-rating.



Back to question list