Biochrom Libra Instruments

Biochrom Libra S4 Visible Spectrophotometer

Biochrom Libra S4 Visible Spectrophotometer

Features

  • Single beam diode array Visible spectrophotometer covering 330-800nm
  • ideal for basic spectrophotometric measurements
  • Easy to use menu driven software
  • Educational experiments and UV/Visible tutorial
  • Grafico PC utility software included  
     

Description
The Biochrom Libra S4 visible spectrophotometer is the ideal instrument for education and QC laboratories. The instrument is compact, lightweight and ergonomically designed for the simpler applications. There is a large display for ease of reading plus a very simple user interface for rapid set up and analysis.

The Libra S4 accepts standard 10mm pathlength glass or disposable cuvettes. An optional test tube adapter set is available for 10, 12 and 16mm plus a range of other accessories. In the event of a spill, the cell holder may be removed for cleaning. The instrument is delivered with a starter pack of disposable cuvettes.

Absorbance, % Transmission and concentration can be measured as well as being able to output absorbance–time plots directly to chart recorder. The instrument is delivered with Grafico PC utility software package and a serial lead, providing the student with the means to capture, print and interpret a wavelength scan from the instrument on a PC; the data for the scan or other results may be easily exported from Grafico to Excel. Grafico also includes an educational tutorial on UV/Visible spectrophotometry and the user manual includes simple experiments for the determination of max, extinction coefficient and natural bandwidth plus the construction of a standard curve and the measurement of stray light.

Ordering Information Order Code
Biochrom Libra S4 Visible Spectrophotometer includes Grafico software, and starter pack of cuvettes 80-5000-00
S1000P Serial printer with serial cable 80-3002-53
Test Tube adaptors (10,12 and 16mm) 80-2117-47
Chart recorder interface cable 80-3003-55

Spare Parts
80-2115-33   Spare Tungsten lamp
80-3001-25   2 rolls printer paper and ribbon for S1000P

Technical Specification shown below. Optional accessories marked *

 
Document Downloads
  File Description File Size  
Adobe PDF Biochrom Libra Range of Spectrophotometers and Colorimeters Brochure
Spectrometers and colorimeters
771 KB Download
Adobe PDF Biochrom Libra Spectrophotometer Order Guide
Cuvettes and accessories for the Biochrom Libra range of spectrophotometers and colorimeters for use in industrial and life science laboratories.
409 KB Download
Adobe PDF Biochrom Libra Range of Spectrophotometers and Colorimeters Brochure French
Product range brochure for the Biochrom Libra range of spectrometers and colorimeters
983 KB Download
Adobe PDF Biochrom Libra Spectrophotometer Order Guide in French
Biochrom Libra accessories and cuvettes for UV/Vis spectrophotometers and colorimeters
705 KB Download
 
Technical Specifications
Specification Data
Type Visible 
Wavelength Range 330 - 800nm 
Wavelength Accuracy (+/-) 2nm 
Wavelength Reproducibility (+/-) 1nm 
Lamp Source Tungsten Halogen 
Beam Single Beam 
Optical System Diode Array 
Bandwidth <7nm 
Detector Diode Array 
Photometric Range -0.300 to 2.500A 
Stray Light <1%T at 340nm 
Zero Stability ±005A/hour near 0A 
Baseline Flatness n/a 
Measurement Type Fixed Wavelength, Abs/%T, concentration (factor), kinetics* 
Cell Compatibility Cuvette 10mm 
Other Sample Containers Test Tube* (10, 12, 16mm) 
Cell Holder Single Cell 
Accessory Options None 
Temperature Control None 
Method Storage None 
Regulatory Compliance CE 
Printer External via Export*, Chart Recorder* 
Data Output Print*, Export to PC* 
Display LCD 
Export/ Data Storage Serial Cable 
Data Interface Serial 
Languages English, manuals in French,German, Italian and Spanish 
Power Requirement 90-265V AC, 50/60Hz, 15VA 
Weight 1.75Kg (3.85lb) 
Dimensions (WxDxH) 31 x 23 x 13cm (12.2 x 9.1 x 5.1 inches) 
Certification CE 
Noise ± 0.002A near 0A and ± 0.020A near 2A at 600nm 
Photometric Accuracy <0.002A at 0A at 546nm 
Photometric Reproducibility <0.002 at 0A and 546nm 
Scan Speed  
Spectral Scanning Yes* (Via Graphico software) 
 
Back to Biochrom Libra Instruments...