Microplate Readers

Biochrom Asys Expert Plus Microplate Reader

Biochrom Asys Expert Plus Microplate Reader
  • Standalone microplate reader supplied with 4 filters (405, 450, 492, 620 nm)
  • Multichannel reading, ideal for labs requiring simple rapid reading of microplates
  • Both onboard and PC software options for flexibility in analytical requirements
  • QC plate and QC software for instrument validation for peace of mind
  • Designed for simple rapid absorbance assays


The Biochrom Asys Expert Plus is a standalone modern design multi-channel microplate reader, for fast plate reading of standard ANSI SBS footprint 96 well, flat, round and v-bottomed plates. The Biochrom Expert Plus has four shaking modes for mixing of solutions and the ability to perform readings at timed intervals ideal for kinetic assays.
Four filters are supplied as standard that cover common absorbance assays. Filters from 400 up to 800 nm can be purchased separately when more specialised assays are required.

Choose the software options to meet your needs
The Biochrom Asys Expert Plus has onboard software for instrument control and data analysis. The onboard software enables instrument control, complete plate setup, plus qualitative, quantitative and kinetics, and a range of curve fit methods, including parabolic, cubic, point to point, spline and 4 parameter fit. Assay formulas can also be setup including validation outliers, elimination, transformation factors and floating cut-offs.
The onboard software can store upto 120 methods and the results of upto 100 plates.

PC-control of the Biochrom Asys Expert Plus for additional analytical capabilities
The standalone Biochrom Expert Plus can be operated in remote mode using PC software. The instrument is supplied with DigiRead which can control the plate reader and allow raw data to be transferred to a Microsoft® Excel spreadsheet for further analysis and Control Plus PC software which simulates the onboard software and allows both methods and raw data to be uploaded and downloaded to and from a PC.


Optional PC software packages:
• KIM is a data reduction package including end-point assay requirements for 96-well plates, and the ability to design bespoke reports


Applications
Designed for ELISA's, enzyme immunoassays, protein quantitation (BCA, Lowry, Bradford), cell proliferation, and cell survival assays, plus cell based assays where shaking is required to keep cells in suspension, enzyme kinetic assays, and high throughput assays where rapid plate reading is required.


Downloadable assay methods
The Biochrom Asys Expert Plus can be downloaded with methods for the Biotrak Assays from GE Healthcare http://www.biochrom.co.uk/user_downloads/?c=19

Technical Support Articles

Biochrom Asys Expert Plus Microtiter Plate Reader Quick Start Guide
Transferring Data from the Biochrom Asys Expert Plus to a PC Using Connect+ Software
Validating the Performance of Biochrom Asys Microplate Readers Using QC Plate and WINQC Software

Ordering Information

Instrument Description Order Code

Biochrom Asys Expert Plus

Standalone Microplate Reader. Wavelength range: 400 - 800 nm, with 4 standard filters (405, 450, 492, 620), including DigiRead and Connect Plus software, operator manual, power cord, RS-232 cable, parallel printer connection cable.

G020150090

Biochrom Asys Expert Plus, including KIM

Standalone Microplate Reader. Wavelength range: 400 - 800 nm, with 4 standard filters (405, 450, 492, 620), including DigiRead and Connect Plus software, operator manual, power cord, RS-232 cable, parallel printer connection cable with KIM data analysis software

G020150010

Accesories
KIM (screening and curve fit)
Software
SK010169
QC Plate QC Plate with WinQC software G010160
Spare lamp For Biochrom Asys Expert 96 and Biochrom Asys Expert Plus SB020007

 
Document Downloads
  File Description File Size  
Adobe PDF Expert Plus CE Certificate
Expert Plus Microplate Reader CE Certificate
22 KB Download
Adobe PDF Biochrom Asys Expert Plus 80-4001-09 Issue 0911 v2
303 KB Download
 
Technical Specifications
Specification Data
Lamp Source Tungsten Halogen 
Measurement Range 0-4.0 OD
Reproducibility/Precision 0.8% and 0.005 OD from 0.10-2.0 OD at 450 nm 
Accuracy 1% from 0.100-2.5 OD 
Standard Filters 340 nm standard with UV model 
Wavelength Selection Filters 
Wavelength Range 400-800 or 340-800nm 
Plate Types flat-bottom, 96- well plate 
Data Output Print*, Export to PC 
Detection System 8 silicon photodiodes 
Reading Speed 5 Seconds 
Software Comprehesive on-board 
Shaking 4 speed standard 
Temperature Control None 
Linearity 0.5% and 0.005 from 0.1-2.5 OD at 492 nm 
Computer Interface RS232 bidirectional 
Optical System Filters 
Power Requirement 90-250V, 50/60 Hz, 80VA 
Validation  
Wavelength Reproducibility (+/-)  
Wavelength Accuracy (+/-)  
Dimensions (WxDxH) 27 x 43 x 24 cm 
Weight 12 kg 
Mode of Operation Local control 
Analysis Software Onboard analysis capabilities 
 