Microplate Readers
EZ Read 2000 Microplate Reader

EZ Read 2000 is the best choice for laboratories that require a microplate reader that can be used for a range of absorbance applications. This versatile microplate reader is capable of measuring samples in many different microplate formats: 12-, 24-, 48- and 96-well plates. The EZ Read 2000 uses tuneable wavelength scanning so that any wavelength can be chosen from 340nm up to 800nm. The EZ Read 2000 is PC-controlled using new, intuitive Galapagos software for data acquisition and analysis.
Galapagos for EZ Read Software
- Galapagos for EZ Read Software makes ‘Click and drag’ approach makes usability a key feature.
- Easy to create new methods.
- Select only the data you want for data reporting
- So easy you won’t need to export your data into Excel.
- Windows 7 & 8 compatibility as well as older Windows OS from XP.
Key Features of EZ Read 2000
- No need for filters! Monochromator optics for measurements from 340 nm to 800 nm enables a wide range of assays to be performed
- Automatic calibration of single channel optical system with self-check and for precision measurements
- Shaking ensures complete mixing of solutions, cells to stay in suspension and colored solutions to be homogenized
- Wavelength scanning for contamination detection, peak absorbance detection and product formation studies
- 12-, 24, 48- and 96-well microplates with standard ANSI SBS footprints can be used for flexible applications
- Real-time reference measurement channel for the improved accuracy
Applications
- Absorbance measurements: ELISA, protein quantitation, cell proliferation and cell survival assays
- Enzyme kinetic measurements requiring automatic readings at timed intervals
- Wavelength scanning to determine sample purity
- Cell-based assays requiring shaking to maintain cells in solution
- Assays measuring the changes of absorbance of NADH at 340nm
- Contamination detection, peak absorbance detection and product formation using scanning across a range of wavelengths
Specification | Data |
---|---|
Lamp Source | Tungsten Halogen |
Measurement Range | 0-3.2 OD |
Reproducibility/Precision | 0.8% and 0.005 OD from 0.10-2.0 OD at 450 nm |
Accuracy | 0.5% and 0.005 OD from 0.1-1.0 OD at 450 nm; 1.0% and 0.010 OD from 1.0-2.0 OD at 450 nm |
Standard Filters | None |
Wavelength Selection | Monochromator |
Wavelength Range | 340-800nm |
Plate Types | 12, 24, 48 and 96-well plates with standard ANSI/SBS footprint |
Data Output | Export to PC, PC Control |
Detection System | 2 silicon diodes: one for measurement, one for reference |
Reading Speed | 35 seconds |
Software | Galapagos Software |
Shaking | Linear shaking in 4 modes |
Temperature Control | None |
Linearity | 0.5% and 0.005 OD from 0.10-1.0 OD at 492 nm 1.0% and 0.010 OD from 1.0-2.0 OD at 492 nm |
Computer Interface | USB 2.0 |
Optical System | Monochromator |
Power Requirement | 90-250 V, 50/60 Hz, 65VA |
Validation | Optional Asys Check plate with WINQC software |
Wavelength Reproducibility (+/-) | n/a |
Wavelength Accuracy (+/-) | 3.0 nm |
Dimensions (WxDxH) | 27 x 43 x 24 cm |
Weight | 10.6 kg |
Mode of Operation | PC Control |